ICU Mednet(TM) Medication Management Suite 6.1 and 6.21, Product Code/List Numbers: (a) 16037-64...

FDA Device Recall #Z-0195-2018 — Class II — October 30, 2017

Recall Summary

Recall Number Z-0195-2018
Classification Class II — Moderate risk
Date Initiated October 30, 2017
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm ICU Medical Inc
Location Lake Forest, IL
Product Type Devices
Quantity 108 installations

Product Description

ICU Mednet(TM) Medication Management Suite 6.1 and 6.21, Product Code/List Numbers: (a) 16037-64-01; MedNet 6.1 b) 16037-64-02; MedNet 6.1 c) 16037-64-03; MedNet 6.1 d) 16037-64-04; MedNet 6.21 The ICU Medical MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and MedNet compatible infusion pumps. The MMS provides trained healthcare professionals with the capability to send, receive, report, and store information from interfaced external systems, and to configure and edit infusion programming parameters. The MMS is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by a trained healthcare professional according to physician s orders.

Reason for Recall

Issue 1: The MedNet Meds 6.1 and 6.21 programs, under certain conditions, can change the piggyback medication entry set settings for existing defined medication entries. This includes (a) whether a medication may be piggybacked and (b) whether an infusion may be interrupted by a piggyback. This only occurs when: 1. Medication entries are built with the same displayed name and diluent amount AND; 2. The piggyback or interruptible settings differ from one CCA to another AND; 3. The drug library is exported then imported into the MedNet Meds program. The changed drug library may then be finalized and downloaded to infusers for use in clinical settings. The changes are made without any notice provided to either the MedNet Meds user (pharmacist) or to the clinician operating the infuser that the piggyback or interruptible settings have been altered. Issue 2: The MedNet Meds 6.1 program, under certain conditions different from those described above in Issue 1, can change the piggyback medication entry set settings for existing defined medication entries. This includes (a) whether a medication may be piggybacked and (b) whether an infusion may be interrupted by a piggyback. This only occurs when: 1. The Master Drug Formulary contains medication entries built with the same displayed name and diluent amounts but one is a Limited rule set and the other is a Full rule set AND; 2. The medication entry built as the Full rule set has a blank medication amount with the medication units in mg AND; 3. Both medication entries have different piggyback options set AND; 4. The drug library is exported then imported into the MedNet Meds program. In this case, the piggyback options will be changed during an import of such drug library.

Distribution Pattern

Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Canada, Australia and Philippines

Lot / Code Information

All MedNet 6.1 and 6.21 installations

Other Recalls from ICU Medical Inc

Recall # Classification Product Date
Z-2449-2024 Class I Plum 360/A+ spare batteries. Item Number: SUB00... May 7, 2024
Z-2448-2024 Class I Replacement Battery List Number SUB0000594 and ... May 7, 2024
Z-2447-2024 Class I Replacement Battery List Number SUB0000864, fou... May 7, 2024
Z-2430-2023 Class II Plum 360 Infusion System, List Numbers: 300100... Jul 13, 2023
Z-1566-2023 Class I Replacement Battery List Number SUB0000594 foun... Mar 22, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.