00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm...
FDA Device Recall #Z-1092-2014 — Class II — January 10, 2014
Recall Summary
| Recall Number | Z-1092-2014 |
| Classification | Class II — Moderate risk |
| Date Initiated | January 10, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Zimmer, Inc. |
| Location | Warsaw, IN |
| Product Type | Devices |
| Quantity | 10 units |
Product Description
00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter 130 mm Stem Length Trabecular Metal Humeral Stem 48 Degree 11mm x 130mm, Sterile, QTY 1, Rx Product Usage: Zimmer Trabecular Metal TM Humeral Stem is a shoulder prosthetic replacement device that is indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint as well as humeral fractures. The humeral stem uses. Trabecular Metal around the proximal part of the stem for biological ingrowth.
Reason for Recall
This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of the minimum required by specification in one of the three zones of the Trabecular Metal. The other two zones exceeded the minimum specification.
Distribution Pattern
Distribution USA in the state of Illinois.
Lot / Code Information
lot 62267250
Other Recalls from Zimmer, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0981-2026 | Class II | Brand Name: Affixus Antegrade Femoral Nailing S... | Dec 2, 2025 |
| Z-0979-2026 | Class II | Brand Name: Affixus Antegrade Femoral Nailing S... | Dec 2, 2025 |
| Z-0973-2026 | Class II | Brand Name: Affixus Antegrade Femoral Nailing S... | Dec 2, 2025 |
| Z-0977-2026 | Class II | Brand Name: Affixus Antegrade Femoral Nailing S... | Dec 2, 2025 |
| Z-0967-2026 | Class II | Brand Name: Affixus Antegrade Femoral Nailing S... | Dec 2, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.