Syngo.plaza, picture archiving and communications system.

FDA Device Recall #Z-1116-2017 — Class II — January 11, 2017

Recall Summary

Recall Number Z-1116-2017
Classification Class II — Moderate risk
Date Initiated January 11, 2017
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Medical Solutions USA, Inc
Location Malvern, PA
Product Type Devices
Quantity 62 systems

Product Description

Syngo.plaza, picture archiving and communications system.

Reason for Recall

Software update for improvements and to resolve several issues

Distribution Pattern

Distributed throughout the United States.

Lot / Code Information

Model Number: 10863171, 10863172, 10863173 Serial Numbers: 100262 100198 100258 100259 100135 100538 100133 100031 100105 100106 100109 100110 100111 100112 100113 100114 100115 100116 100117 100119 100120 100416 100461 100245 100447 100364 100298 100299 100386 100139 100027 100440 100441 100559 100025 100033 100196 100189 100190 100191 100192 100195 100186 100232 100450 100449 100256 100199 100188 100194 100198 100029 100253 100203 100204 100132 100139 100434 100476 100495 100491 100206

Other Recalls from Siemens Medical Solutions USA, Inc

Recall # Classification Product Date
Z-1175-2026 Class II Interventional Fluoroscopic X-Ray System: LUMIN... Dec 29, 2025
Z-1174-2026 Class II Interventional Fluoroscopic X-Ray System: LUMIN... Dec 29, 2025
Z-1241-2026 Class II NAEOTOM Alpha.Prime Software applications: s... Dec 19, 2025
Z-1240-2026 Class II NAEOTOM Alpha Software applications: syngo.C... Dec 19, 2025
Z-1239-2026 Class II SOMATOM X.ceed Software applications: syngo.... Dec 19, 2025

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.