Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8...
FDA Recall #Z-0492-2022 — Class II — December 20, 2021
Product Description
Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F, and INFX-8000H.
Reason for Recall
Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel.
Recalling Firm
Canon Medical System, USA, INC. — Tustin, CA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
8
Distribution
US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV.
Code Information
SID: 30039059, Alphenix System Serial Number: BGC20Y2010, & Flat Panel Detector Serial Number: C1B20Y2215 SID: 30030612, Alphenix System Serial Number: B8D20Z2001, & Flat Panel Detector Serial Number: C1B20Z2218; SID: 30030612, Alphenix System Serial Number: B8D20Z2001, & Flat Panel Detector Serial Number: C1B20Z2219; SID: 30046247, Alphenix System Serial Number: B1B20Z2001, & Flat Panel Detector Serial Number: C1B20Z2221; SID: 30064041, Alphenix System Serial Number: BGD2172011, & Flat Panel Detector Serial Number: C1B2172234; SID: 30065145, Alphenix System Serial Number: B1E2172002, & Flat Panel Detector Serial Number: C1B2172235; SID: 30058149, Alphenix System Serial Number: B1E2152001, & Flat Panel Detector Serial Number: C1B2152230; SID: 30002089, Alphenix System Serial Number: BHC20Z2003, & Flat Panel Detector Serial Number: C1B20Z2217.
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated