Baxter Spectrum Wireless Battery Modules A/B/G/N, product code 35223, used with either of the fol...

FDA Device Recall #Z-2581-2024 — Class II — July 17, 2024

Recall Summary

Recall Number Z-2581-2024
Classification Class II — Moderate risk
Date Initiated July 17, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Baxter Healthcare Corporation
Location Deerfield, IL
Product Type Devices
Quantity 17,377 units

Product Description

Baxter Spectrum Wireless Battery Modules A/B/G/N, product code 35223, used with either of the following: 35700BAX2 (Spectrum V8 infusion pump) or 3570009 (Spectrum IQ infusion pump) Baxter SIGMA Spectrum V8 Infusion Pump

Reason for Recall

The Spectrum WBM may fail to auto-document infusion status information back to the hospital's Electronic Medical Record (EMR) system. This software issue only impacts customers where the Spectrum V8 or Spectrum IQ infusion system is integrated with the hospital s EMR system.

Distribution Pattern

US, Nationwide

Lot / Code Information

Pump GTIN 00085412498683; All serial numbers with radio software versions 22D28, 22D29, and 22D30

Other Recalls from Baxter Healthcare Corporation

Recall # Classification Product Date
Z-1573-2026 Class II Brand Name: Welch Allyn, Inc. Product Name: We... Feb 18, 2026
Z-1368-2026 Class II Brand Name: Spectrum IQ Product Name: Infusion... Jan 13, 2026
Z-1121-2026 Class II SIGMA Spectrum Infusion Pump, Model/Catalog Num... Nov 28, 2025
Z-0847-2026 Class II Welch Allyn INF BAG, Thigh 1-Tube, Model/Catalo... Oct 28, 2025
Z-0233-2026 Class II Oral Probe (Product code 02893-000), accessory ... Sep 17, 2025

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.