Microdot Xtra blood glucose test strips

FDA Device Recall #Z-2357-2019 — Class II — July 26, 2018

Recall Summary

Recall Number Z-2357-2019
Classification Class II — Moderate risk
Date Initiated July 26, 2018
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Cambridge Sensors Limited
Location Huntingdon, N/A
Product Type Devices
Quantity 17,784 (USA)

Product Description

Microdot Xtra blood glucose test strips

Reason for Recall

The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.

Distribution Pattern

Illinois

Lot / Code Information

LOT NUMBERS 7091101, 7091201, 7091801, 7091902, 7100402, 7100403, 7101002, 7101602, 7101702, 7102403, 7103002, 7110602, 7110703, and 7112702.

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.