Aplio 500/400/300 Diagnostic Ultrasound System (TUS-500/400/300). Device Listing Number: Dl53118...

FDA Recall #Z-1104-2013 — Class II — December 11, 2012

Recall #Z-1104-2013 Date: December 11, 2012 Classification: Class II Status: Terminated

Product Description

Aplio 500/400/300 Diagnostic Ultrasound System (TUS-500/400/300). Device Listing Number: Dl53118. Diagnostic Ultrasound system.

Reason for Recall

Toshiba Medical Systems has received reports of startup issues with the Toshiba TUS-500 and TUS-300 ultrasound systems. The startup issue is predominant when attempting to bring the system up from standby mode when the system is used remotely. It has been determined that this may potentially cause damage to the system.

Recalling Firm

Toshiba American Medical Systems Inc — Tustin, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

233 units

Distribution

Nationwide Distribution.

Code Information

Item numbers: TUS-A300/Wl and TUS-A500/Wl

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated