PSA Assays for use in the IMMULITE/IMMULITE 1000 Systems Analyzers, Catalog Numbers LKPS1, LKPS5 ...

FDA Device Recall #Z-2047-2013 — Class II — June 26, 2013

Recall Summary

Recall Number Z-2047-2013
Classification Class II — Moderate risk
Date Initiated June 26, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Healthcare Diagnostics
Location Tarrytown, NY
Product Type Devices
Quantity 24,361 kits

Product Description

PSA Assays for use in the IMMULITE/IMMULITE 1000 Systems Analyzers, Catalog Numbers LKPS1, LKPS5 and LKPTS1(D); and in the IMMULITE 2000/2000XPi Systems Analyzers, Catalog Numbers L2KPS2, L2KPS6, L2KPTS2(D) and L2KPTS6(D). IVD Immunoassay kits in the following sizes: 100, 200, 500 and 600 test kits; For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 and the IMMULITE 2000/ 2000 XPi Systems Analyzers. PMA # P930027/ Device Listing # D002374. Product Usage: For the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients. For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 and the IMMULITE 2000/ 2000 XPI Systems Analyzers

Reason for Recall

Customers complained of a positive bias in patient results on the IMMULITE/ IMMULITE 1000 and the IMMULITE 2000/ 2000 XPi PSA assay when compared against the IMMULITE /IMMULITE 1000 3rd Generation PSA assay. Siemens has confirmed an overall average positive bias of approximately 20% to 23% across the assay range relative to the WHO 96/670 with the IMMMULITE Systems PSA Assay. This positive bias is observed in patient values and the Siemens Tumor Marker controls (TMCO). Other commercially available controls may show this bias.

Distribution Pattern

Worldwide Distribution - USA Nationwide including Puerto Rico and the countries of: Albania, Afghanistan, Algeria, Argentina, Austria, Bangladesh, Belgium, Bolivia, Bosnia- Herzegovinia, Brazil, Bulgaria, Camaroon, Canada, Canary Islands, Chile, China, Colombia, Coast Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Honduras, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Malta, Mexico, Morocco, Nicaragua, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, Spain, South Africa, Rep. of Korea (S), Sweden, Switzerland, Syria, Taiwan, Thailand, The Netherlands, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, Uruguay, White Russia, Venezuela, and Vietnam.

Lot / Code Information

Product Code/Lot # LKPS1 424, Expiration Date 2013-08-31; Product Code/Lot # LKPS1 425, Expiration Date 2013-10-31; Product Code/Lot # LKPS1 426, Expiration Date 2013-10-31; Product Code/Lot # LKPS5 424, Expiration Date 2013-08-31; Product Code/Lot # LKPS5 425, Expiration Date 2013-10-31; Product Code/Lot # LKPS5 426, Expiration Date 2013-10-31; Product Code/Lot # LKPTS1(D) D0104, Expiration Date 2013-08-31; Product Code/Lot # L2KPS2 379, Expiration Date 2013-07-31; Product Code/Lot # L2KPS2 380, Expiration Date 2013-08-31; Product Code/Lot # L2KPS2 381, Expiration Date 2013-10-31; Product Code/Lot # L2KPS2 382, Expiration Date 2013-10-31; Product Code/Lot # L2KPS2 383, Expiration Date 2013-10-31; Product Code/Lot # L2KPS2 384, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 379, Expiration Date 2013-07-31; Product Code/Lot # L2KPS6 380, Expiration Date 2013-08-31; Product Code/Lot # L2KPS6 381, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 382, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 383, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 384, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS2(D) D107, Expiration Date 2013-07-31; Product Code/Lot # L2KPTS2(D) D108, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS2(D) D109, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS6(D) D107, Expiration Date 2013-07-31; Product Code/Lot # L2KPTS6(D) D108, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS6(D) D109, Expiration Date 2013-10-31.

Other Recalls from Siemens Healthcare Diagnostics

Recall # Classification Product Date
Z-0827-2018 Class II BC Thrombin Reagent Kit Lot Number 46751; UDI -... Oct 17, 2017
Z-0122-2015 Class II IMMULITE /IMMULITE 1000 Systems COR Cortisol, R... Sep 19, 2014
Z-2715-2014 Class II ADVIA Centaur Immunoassay System (including ref... Aug 19, 2014
Z-2716-2014 Class II ADVIA Centaur XP Immunoassay System (including ... Aug 19, 2014
Z-2662-2014 Class II ADVIA Centaur Immunoassay System, Catalog Numbe... Aug 13, 2014

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.