Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes...
FDA Recall #Z-1252-2017 — Class II — December 12, 2016
Product Description
Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney.
Reason for Recall
Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section during surgical procedures. Some of these complaints are associated with tissue trauma, including perforation, and insertion tubes which were stuck inside the patient and had to be surgically removed.
Recalling Firm
Olympus Corporation of the Americas — Center Valley, PA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
1,627 units (769 URF-V2 and 858 URF-V2R)
Distribution
Worldwide distribution-US Nationwide and country of: Canada.
Code Information
All serial numbers
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated