Stryker iBur 3.0mm Precision Match Head, Distal Bend- intended to cut bone in the following man...

FDA Recall #Z-2731-2024 — Class II — July 24, 2024

Recall #Z-2731-2024 Date: July 24, 2024 Classification: Class II Status: Completed

Product Description

Stryker iBur 3.0mm Precision Match Head, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-107-530

Reason for Recall

Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.

Recalling Firm

Stryker Corporation — Portage, MI

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

3673 units

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.

Code Information

GTIN: 07613327501193 Lot Numbers: 22228017 22257017 22269017 22276017 23085017 23095017 23157017 23157027 23179017 23179027 23198017 23213017 23221017 23299017 23345017 23345027

Status

Completed

Voluntary / Mandated

Voluntary: Firm initiated