Dynavox T15 Speech Generating Device ( including Batteries)

FDA Recall #Z-1305-2018 — Class II — December 14, 2017

Recall #Z-1305-2018 Date: December 14, 2017 Classification: Class II Status: Terminated

Product Description

Dynavox T15 Speech Generating Device ( including Batteries)

Reason for Recall

The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.

Recalling Firm

Dynavox Systems Llc — Pittsburgh, PA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

307

Distribution

Nationally

Code Information

TL1405 TL1407 Part Numbers: T15: 750423 Date Codes are part of the device s unique Serial Number and represent the year and month the device was manufactured.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated