Smith & Nephew Taylor Spatial Frame Website software, Catalog Number 71070401. A web-based and m...

FDA Recall #Z-0409-2018 — Class II — December 13, 2017

Recall #Z-0409-2018 Date: December 13, 2017 Classification: Class II Status: Terminated

Product Description

Smith & Nephew Taylor Spatial Frame Website software, Catalog Number 71070401. A web-based and mobile application software component of a multilateral external fixation system that is intended for fracture fixation, fixation of long bones and for joint fusions and limb lengthening or deformity corrections which involve cutting of the bone.

Reason for Recall

In rare circumstances, when surgeons use the software to plan the strut adjustment, an inverted image of the frame is generated by the software. The frame will appear inverted in the preview images on the Strut Settings tab.

Recalling Firm

Smith & Nephew, Inc. — Memphis, TN

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

N/A

Distribution

Worldwide

Code Information

Version 5.2.5 and 5.2.6

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated