VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic u...

FDA Device Recall #Z-1469-2020 — Class II — August 12, 2018

Recall Summary

Recall Number Z-1469-2020
Classification Class II — Moderate risk
Date Initiated August 12, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm The Binding Site Group, Ltd.
Location Birmingham
Product Type Devices
Quantity 462 kits

Product Description

VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.

Reason for Recall

A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of M.A., C.A., MO., NC, MN, VA, WA, and IL and country of Canada.

Lot / Code Information

UDI: 05051700006767. Lot numbers 413170 and 435314.

Other Recalls from The Binding Site Group, Ltd.

Recall # Classification Product Date
Z-0274-2022 Class II Freelite Human Lambda Free Kit (for use on the ... Oct 15, 2021
Z-2030-2021 Class II FREELITE Human Kappa Free Kit for use on the Ro... May 20, 2021
Z-0943-2021 Class II Optilite Freelite Mx Kappa Free Kit REF LK016.... Nov 18, 2020
Z-2071-2020 Class II The Optilite Clinical Chemistry Analyzer. IVD. ... Apr 9, 2020
Z-1500-2020 Class II Rheumatoid Factor (RF) Kit for use on SPAPLUS, ... Nov 6, 2019

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.