STAR ActiveTrak Excimer Laser System, Model: STAR S3, Catalog Number: 0030-2450, Product is manuf...

FDA Device Recall #Z-0047-2014 — Class II — September 16, 2013

Recall Summary

Recall Number Z-0047-2014
Classification Class II — Moderate risk
Date Initiated September 16, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Abbott Medical Optics, Inc.
Location Milpitas, CA
Product Type Devices
Quantity 834 units in total

Product Description

STAR ActiveTrak Excimer Laser System, Model: STAR S3, Catalog Number: 0030-2450, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.

Reason for Recall

This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).

Distribution Pattern

Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.

Lot / Code Information

All units of these models: STAR S2 Excimer Laser System; STAR S3 ActiveTrak; STAR S4 with Variable Spot Scanning; and STAR S41R; There are no expiration dates for the STAR Excimer Laser Systems.

Other Recalls from Abbott Medical Optics, Inc.

Recall # Classification Product Date
Z-1515-2015 Class II OptiMedica Catalys Precision Laser System; Mode... Feb 20, 2015
Z-0048-2014 Class II STAR Excimer Laser System with Variable Spot Sc... Sep 16, 2013
Z-0046-2014 Class II STAR Excimer Laser System, Model: STAR S2, Cata... Sep 16, 2013
Z-0049-2014 Class II STAR Excimer Laser System, Model STAR S4 IR, Ca... Sep 16, 2013
Z-1540-2013 Class II STAR S4 with Variable Spot Scanning Excimer Sy... May 28, 2013

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.