Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 ...

FDA Recall #Z-2718-2024 — Class II — June 2, 2024

Recall #Z-2718-2024 Date: June 2, 2024 Classification: Class II Status: Ongoing

Product Description

Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.

Reason for Recall

Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.

Recalling Firm

Treace Medical Concepts, Inc. — Ponte Vedra, FL

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

40 kits containing 40 plates

Distribution

US Nationwide distribution in the states of CO, FL, MD, MI, NC, SC, TX, VA.

Code Information

UDI: (01)00810111222464(17)290424(10)300542404

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated