VERIFY Dual Species Self-Contained Biological Indicators, 100 per box Item Number: S3061 Inte...

FDA Recall #Z-0779-2023 — Class II — December 12, 2022

Recall #Z-0779-2023 Date: December 12, 2022 Classification: Class II Status: Ongoing

Product Description

VERIFY Dual Species Self-Contained Biological Indicators, 100 per box Item Number: S3061 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes

Reason for Recall

Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.

Recalling Firm

Steris Corporation — Mentor, OH

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

857 boxes

Distribution

Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA

Code Information

UDI: 10724995023833, 50724995023824 lot #230613

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.

Back to All Device Recalls