ARTIS pheno- angiography systems developed for single and biplane diagnostic imaging and interven...
FDA Device Recall #Z-2029-2021 — Class II — May 27, 2021
Recall Summary
| Recall Number | Z-2029-2021 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 27, 2021 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Siemens Medical Solutions USA, Inc |
| Location | Malvern, PA |
| Product Type | Devices |
| Quantity | 116 units |
Product Description
ARTIS pheno- angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000
Reason for Recall
Software error VE10B -In case of a fault within the drivetrain of an axis and while performing a stepped peripheral imaging (PERISTEPPING/PERIVISION), the C-arm may leave its intended travel path by up to 22 cm (9") result in canceling clinical treatment or to continue treatment on an alternative system
Distribution Pattern
US Nationwide
Lot / Code Information
Software VE10B. Serial Numbers: 164138 164159 164033 164287 164282 164210 164183 164184 164311 164142 164071 164328 164034 164318 164244 164242 164306 164155 164189 164271 164277 164173 164292 164231 164057 164054 164055 164080 164132 164329 164072 164088 164190 164333 164025 164026 164027 164338 164094 164241 164326 164267 164186 164068 164121 164157 164019 164152 164161 164013 164069 164213 164073 164257 164293 164263 164298 164060 164083 164113 164021 164066 164124 164125 164279 164075 164074 164084 164086 164098 164028 164153 164330 164295 164234 164321 164270 164275 164315 164163 164158 164053 164171 164134 164253 164188 164063 164064 164299 164300 164105 164160 164024 164049 164240 164227 164228 164018 164022 164250 164061 164187 164115 164322 164269 164303 164204 164205 164211 164212 164216 164196 164168 164319 164201 164202
Other Recalls from Siemens Medical Solutions USA, Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1175-2026 | Class II | Interventional Fluoroscopic X-Ray System: LUMIN... | Dec 29, 2025 |
| Z-1174-2026 | Class II | Interventional Fluoroscopic X-Ray System: LUMIN... | Dec 29, 2025 |
| Z-1241-2026 | Class II | NAEOTOM Alpha.Prime Software applications: s... | Dec 19, 2025 |
| Z-1240-2026 | Class II | NAEOTOM Alpha Software applications: syngo.C... | Dec 19, 2025 |
| Z-1239-2026 | Class II | SOMATOM X.ceed Software applications: syngo.... | Dec 19, 2025 |
Frequently Asked Questions
A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.