Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiov...

FDA Device Recall #Z-0416-2022 — Class II — November 19, 2021

Recall Summary

Recall Number Z-0416-2022
Classification Class II — Moderate risk
Date Initiated November 19, 2021
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Philips North America Llc
Location Cambridge, MA
Product Type Devices
Quantity USA: 5 systems R.O.W.: 15 systems

Product Description

Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. Model Number: 722005, 722006, 722008, 722010, 722012, 722013, 722025, 722026, 722027, 722028.

Reason for Recall

Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of DC, GA, MA, TX, VA and the countries of Dominican Republic, INDIA, Japan, Poland, Spain, Taiwan.

Lot / Code Information

US Serial Number/UDI: 1 (01)00884838054202(21)1; 527 (01)00884838059054(21)527; 408 (01)00884838054202(21)408; 306 (01)00884838054202(21)306; 61 (01)00884838059030(21)61 OUS: Model Product S/N UDI: 722006 Allura Xper FD20 1997 (01)00884838059191(10)1997; 722010 Allura Xper FD10 1177 (01)00884838059030(21)1177; 722010 Allura Xper FD10 1178 (01)00884838059030(21)1178; 722013 Allura Xper FD20 Biplane 544;; 722010 Allura Xper FD10 514 (01)00884838059030(21)514; 722013 Allura Xper FD20 Biplane 271; 722025 Allura Xper FD20 Biplane OR Table 6; 722005 Allura Xper FD10/10 519; 722008 Allura Xper FD20 Biplane 531; 722012 Allura Xper FD20 1901 (01)00884838059054(21)1901; 722012 Allura Xper FD20 2505 (01)00884838059054(21)2505; 722013 Allura Xper FD20 Biplane 137; 722026 Allura Xper FD10 439 (01)00884838054189(21)439; 722134 Field ext. Xper vascular systems R7.6 98736; 722027 Allura Xper FD10/10 172 (01)00884838054196(21)172;

Other Recalls from Philips North America Llc

Recall # Classification Product Date
Z-1316-2026 Class II IQon Spectral CT; Product Code (REF): 728332; Jan 7, 2026
Z-1318-2026 Class II Spectral CT 7500 on Rails; Product Code (REF):... Jan 7, 2026
Z-1317-2026 Class II Spectral CT; Product Code (REF): 728333; Jan 7, 2026
Z-1315-2026 Class II Brilliance iCT; Product Code (REF): 728306; Jan 7, 2026
Z-1033-2026 Class II Philips Smart-hopping 2.0 AP 1.4 GHz. Patient M... Dec 12, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.