epoc NXS Host; Siemens Material Number (SMN): 11413506 (CA); Software Version Number: 4.14.9. ...

FDA Device Recall #Z-0737-2025 — Class II — November 22, 2024

Recall Summary

Recall Number Z-0737-2025
Classification Class II — Moderate risk
Date Initiated November 22, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Healthcare Diagnostics, Inc.
Location East Walpole, MA
Product Type Devices
Quantity 320 units

Product Description

epoc NXS Host; Siemens Material Number (SMN): 11413506 (CA); Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

Reason for Recall

Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.

Distribution Pattern

Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kuwait Latvia Lithuania Luxembourg Macau Madagascar Malaysia Mexico Monaco Mongolia Morocco Myanmar Netherlands New Caledonia New Zealand Norway Paraguay Peru Philippines Poland Portugal Republic Korea Romania Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Swaziland Sweden Switzerland Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vietnam

Lot / Code Information

Siemens Material Number: 11413506; UDI-DI: 00630414605821; Software Version Number: 4.14.9.

Other Recalls from Siemens Healthcare Diagnostics, Inc.

Recall # Classification Product Date
Z-1484-2026 Class II Atellica CH Urine Albumin (UAlb). Material Numb... Jan 20, 2026
Z-0491-2026 Class II IMMULITE 2000 Oak Mix Specific Allergen; Sieme... Oct 6, 2025
Z-0492-2026 Class II 3gAllergy Specific IgE Universal Kit; Kit Comp... Oct 6, 2025
Z-0493-2026 Class II Food Panel 3 Specific Allergen; Siemens Materi... Oct 6, 2025
Z-0366-2026 Class II Assay: IMMULITE 2000 Intact PTH; Test Code: iP... Sep 22, 2025

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.