Artis Q, and Artis Q.zen ceiling systems; model numbers: 10848281; 10848354: The Artis family is ...

FDA Recall #Z-0972-2015 — Class II — December 3, 2014

Recall #Z-0972-2015 Date: December 3, 2014 Classification: Class II Status: Terminated

Product Description

Artis Q, and Artis Q.zen ceiling systems; model numbers: 10848281; 10848354: The Artis family is dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.

Reason for Recall

Suboptimal routing of the cable may result in increased wear over time. Without additional measures, damage of a cable may result in limited functionality and in rare cases, failure of a system. An ongoing procedure could be terminated in rare cases.

Recalling Firm

Siemens Medical Solutions USA, Inc — Malvern, PA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

50

Distribution

Nationwide Distribution

Code Information

model numbers: 10848281; 10848354

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.