CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mo...
FDA Device Recall #Z-1281-2016 — Class II — February 16, 2016
Recall Summary
| Recall Number | Z-1281-2016 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 16, 2016 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Siemens Medical Solutions USA, Inc |
| Location | Malvern, PA |
| Product Type | Devices |
| Quantity | 4 units |
Product Description
CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
Reason for Recall
Siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is only present when a service technician ignores several steps listed in the service manual as precautionary actions during device maintenance.
Distribution Pattern
US Nationwide Distributed in the states of AR, CA, NE and UT
Lot / Code Information
Model Number: 10308191
Other Recalls from Siemens Medical Solutions USA, Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1175-2026 | Class II | Interventional Fluoroscopic X-Ray System: LUMIN... | Dec 29, 2025 |
| Z-1174-2026 | Class II | Interventional Fluoroscopic X-Ray System: LUMIN... | Dec 29, 2025 |
| Z-1241-2026 | Class II | NAEOTOM Alpha.Prime Software applications: s... | Dec 19, 2025 |
| Z-1240-2026 | Class II | NAEOTOM Alpha Software applications: syngo.C... | Dec 19, 2025 |
| Z-1239-2026 | Class II | SOMATOM X.ceed Software applications: syngo.... | Dec 19, 2025 |
Frequently Asked Questions
Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.