Artis zee multi, Model Number 10094139

FDA Device Recall #Z-2030-2019 — Class II — June 12, 2019

Recall Summary

Recall Number Z-2030-2019
Classification Class II — Moderate risk
Date Initiated June 12, 2019
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Medical Solutions USA, Inc
Location Malvern, PA
Product Type Devices
Quantity 89

Product Description

Artis zee multi, Model Number 10094139

Reason for Recall

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Distribution Pattern

US Nationwide Distribution

Lot / Code Information

Serial Numbers: 157435 157509 157660 157710 157458 130017 157698 157532 157449 157657 157136 157690 157616 157560 157436 157677 157528 157503 157615 157448 157165 157700 157602 157168 157642 157213 157164 157559 157156 157524 157588 157750 157579 157573 157650 157671 157554 157555 157521 157696 157488 157644 157648 157169 157668 157508 157510 157601 157139 158058 157536 157625 157600 157604 157565 157634 157635 157146 157583 157476 157497 157570 157658 157505 157594 157706 157586 157674 157414 157215 157692 157707 157507 157708 157544 157184 157659 157661 157531 157409 157688 157702 157704 157466 157442 157667 157438 157440 157645

Other Recalls from Siemens Medical Solutions USA, Inc

Recall # Classification Product Date
Z-1175-2026 Class II Interventional Fluoroscopic X-Ray System: LUMIN... Dec 29, 2025
Z-1174-2026 Class II Interventional Fluoroscopic X-Ray System: LUMIN... Dec 29, 2025
Z-1241-2026 Class II NAEOTOM Alpha.Prime Software applications: s... Dec 19, 2025
Z-1240-2026 Class II NAEOTOM Alpha Software applications: syngo.C... Dec 19, 2025
Z-1239-2026 Class II SOMATOM X.ceed Software applications: syngo.... Dec 19, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.