Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number ...
FDA Recall #Z-1574-2023 — Class II — November 19, 2021
Product Description
Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal surgery
Reason for Recall
Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.
Recalling Firm
Medtronic Sofamor Danek USA Inc — Memphis, TN
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
4 units
Distribution
US: IN, CT, NE
Code Information
UDI/DI (GTIN) 00763000401535, Lot Numbers EM21D006, EM21D007; Loaner Kit Lot numbers: 0072, 0073, 0077, 0098, 0103
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.