Bio-Rad, REF: 12005660, QXDx BCR-ABL %IS Kit

FDA Device Recall #Z-1358-2024 — Class II — February 15, 2024

Recall Summary

Recall Number Z-1358-2024
Classification Class II — Moderate risk
Date Initiated February 15, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Bio-Rad Laboratories
Location Pleasanton, CA
Product Type Devices
Quantity 33 units

Product Description

Bio-Rad, REF: 12005660, QXDx BCR-ABL %IS Kit

Reason for Recall

Due to incorrect calibrations and controls, their is a potential of incorrect reporting of lower %IS and molecular response values.

Distribution Pattern

US Nationwide distribution in the states of PA, SD, TX, FL, MN.

Lot / Code Information

Lot # 64562656 & 64571846/ UDI: 03610521162834

Other Recalls from Bio-Rad Laboratories

Recall # Classification Product Date
Z-0032-2014 Class III FSE Image Upgrade Kit, Catalog # 92019C This... Aug 20, 2013
Z-2398-2012 Class II The Bio-Rad MONOLISA Anti-HBs EIA is for In Vit... Jul 27, 2012

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.