Tina-quant Complement C4 ver.2, Catalog 05991994190
FDA Device Recall #Z-0103-2021 — Class III — August 27, 2020
Recall Summary
| Recall Number | Z-0103-2021 |
| Classification | Class III — Low risk |
| Date Initiated | August 27, 2020 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Roche Diagnostics Operations, Inc. |
| Location | Indianapolis, IN |
| Product Type | Devices |
| Quantity | 180 units |
Product Description
Tina-quant Complement C4 ver.2, Catalog 05991994190
Reason for Recall
Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.
Distribution Pattern
Domestic distribution nationwide.
Lot / Code Information
lot 36870301
Other Recalls from Roche Diagnostics Operations, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1476-2026 | Class II | cobas pro integrated solutions with cobas c 503... | Jan 20, 2026 |
| Z-1009-2026 | Class II | The Elecsys Anti-TSHR immunoassay is a three-st... | Nov 21, 2025 |
| Z-0489-2025 | Class II | cobas HCYS, Homocysteine Enzymatic Assay, Mater... | Oct 17, 2024 |
| Z-3187-2024 | Class II | Creatine Kinase (CK) used on cobas c 311/501/50... | Aug 15, 2024 |
| Z-0352-2024 | Class II | Tina-quant D-Dimer Test System, Fibrinogen/Fibr... | Oct 3, 2023 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.