Tina-quant Complement C4 ver.2, Catalog 05991994190

FDA Device Recall #Z-0103-2021 — Class III — August 27, 2020

Recall Summary

Recall Number Z-0103-2021
Classification Class III — Low risk
Date Initiated August 27, 2020
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Roche Diagnostics Operations, Inc.
Location Indianapolis, IN
Product Type Devices
Quantity 180 units

Product Description

Tina-quant Complement C4 ver.2, Catalog 05991994190

Reason for Recall

Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Distribution Pattern

Domestic distribution nationwide.

Lot / Code Information

lot 36870301

Other Recalls from Roche Diagnostics Operations, Inc.

Recall # Classification Product Date
Z-1476-2026 Class II cobas pro integrated solutions with cobas c 503... Jan 20, 2026
Z-1009-2026 Class II The Elecsys Anti-TSHR immunoassay is a three-st... Nov 21, 2025
Z-0489-2025 Class II cobas HCYS, Homocysteine Enzymatic Assay, Mater... Oct 17, 2024
Z-3187-2024 Class II Creatine Kinase (CK) used on cobas c 311/501/50... Aug 15, 2024
Z-0352-2024 Class II Tina-quant D-Dimer Test System, Fibrinogen/Fibr... Oct 3, 2023

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.