IRON assay, Reference Numbers 6K95-41 and 6K95-30

FDA Device Recall #Z-0325-2024 — Class II — September 29, 2023

Recall Summary

Recall Number Z-0325-2024
Classification Class II — Moderate risk
Date Initiated September 29, 2023
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Sentinel CH SpA
Location Milan, N/A
Product Type Devices
Quantity 4394 kits

Product Description

IRON assay, Reference Numbers 6K95-41 and 6K95-30

Reason for Recall

A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.

Distribution Pattern

Worldwide - US Nationwide distribution.

Lot / Code Information

Reference Number 6K95-41: UDI-DI: 00380740010928; Lot Number: 30610Y600; Reference Number 6K95-30: UDI-DI: 00380740010911; Lot Numbers: 21058Y600, 21101Y600, 30918Y600

Other Recalls from Sentinel CH SpA

Recall # Classification Product Date
Z-0326-2024 Class II Alinity c Iron Reagent, Reference Number 08P3920 Sep 29, 2023
Z-0033-2022 Class III Alinity c CRP Vario Cardiac High Sensitivity Ca... Jul 29, 2021
Z-2012-2021 Class III ACE Control Set - IVD Control set for the quant... Apr 30, 2021
Z-2551-2019 Class II MULTIGENT Lithium-intended for the quantitation... Jun 28, 2019
Z-2552-2019 Class II Alinity c Lithium Reagent Kit- intended for the... Jun 28, 2019

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.