MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Fi...

FDA Recall #Z-1024-2026 — Class II — December 10, 2025

Recall #Z-1024-2026 Date: December 10, 2025 Classification: Class II Status: Ongoing

Product Description

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Reason for Recall

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Recalling Firm

REPLIGEN CORPORATION — Marlborough, MA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

33

Distribution

U.S.

Code Information

PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Status

Ongoing

Voluntary / Mandated

FDA Mandated