Instrument Kit for use with OsteoSponge SC grafts. The Instrument Kit is used to insert OsteoSpo...
FDA Device Recall #Z-1396-2015 — Class II — March 9, 2015
Recall Summary
| Recall Number | Z-1396-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | March 9, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Bacterin International, Inc. |
| Location | Belgrade, MT |
| Product Type | Devices |
| Quantity | 63 units |
Product Description
Instrument Kit for use with OsteoSponge SC grafts. The Instrument Kit is used to insert OsteoSponge SC grafts to fill bone defects in the subchondral region of articulating joints. The Instrument Kit is used to establish and maintain a port of entry for arthroscopic instruments or for an open procedure. The kit is also used to insert the OsteoSponge¿ SC graft during surgery. Such procedures include, but are not limited to, subchondral bone repair. Affected Part numbers are 6508 and 6511.
Reason for Recall
Instrument Kits for OsteoSponge SC grafts, OsteoLock Facet Stabilization Dowel, and BacFast HD Facet Dowel have incorrect sterilization parameters in the IFU and on the lid of the kits. Sterilization parameters for the Steam Gravity 121 degree C (250 F) cycle were not validated and the dry time for the Steam Pre-Vacuum 132 degree C (270 F) cycle was omitted.
Distribution Pattern
US Distribution in states of: Alabama, Arizona, Arkansas, Colorado, Florida, Georgia, Illinois, Indiana, Louisiana, Maryland, Michigan, Minnesota, Mississippi, Missouri, Nebraska, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Puerto Rico, Texas, and Virginia.
Lot / Code Information
SERIAL NUMBERS: 004, 005, 006, 009, 010, 011, 012, 018, 019, 020, 021, 022, 023, 024, 025, 026, 029, 30, 033, 036, 038, 040, 045, 048, 060, 063, 064, 065, 066, 067, 068, 069, 070, 071, 072, 073, 074, 075, 076, 077, 079, 081, 082, 083, 084, 085, 086, 087, 088, 089, 090, 091, 092, 093, 094, 095, 096, 097, 098, 099, 100, 101, and 102.
Other Recalls from Bacterin International, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1680-2014 | Class II | Wound Drain - Label reads in part "Elutia ***Hu... | May 15, 2014 |
| Z-1138-2013 | Class II | OsteoSelect DBM Putty Product Usage: Ortho... | Feb 21, 2013 |
| Z-0936-2013 | Class II | Bacterin Elutia, Hubless Coated Silicone Surgic... | Jan 28, 2013 |
| Z-0800-2013 | Class II | Bacterin International, Inc., Osteoselect Demin... | Apr 18, 2012 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.