Perioscopy Sheath: Part number: 1008; The Perioscopy Sheath is a single-use sterile compo...
FDA Device Recall #Z-2530-2016 — Class II — June 24, 2016
Recall Summary
| Recall Number | Z-2530-2016 |
| Classification | Class II — Moderate risk |
| Date Initiated | June 24, 2016 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Danville Materials, Inc. |
| Location | San Ramon, CA |
| Product Type | Devices |
| Quantity | ~ 2123 units |
Product Description
Perioscopy Sheath: Part number: 1008; The Perioscopy Sheath is a single-use sterile component used with the Perioscopy System, which is a medical device that features miniaturized digital video technology, and enables the healthcare professional to see magnified details of the tooth anatomy and deposits below the gum line (i.e. calculus root fractures, restoration margins, tooth decay).
Reason for Recall
The Perioscopy Sheaths are currently provided with a label that indicates the Sheath has a 3 year shelf-life/expiry date. Danville Materials has become aware that validation data are not available to support the 3-year shelf-life/expiration dating claim.
Distribution Pattern
US and New Zealand, Australia, Canada, Japan, Israel, United Kingdom, Italy, Mexico.
Lot / Code Information
Part number 1008: Lot numbers: 1133; 1135; 1136; 1137; 1138; 1163; 1171; 1172; 1173; 1174; 1175; 1218; 1240; 1243; 1258; 1264; 1264-1; 1275; 1276; 1277; 1311; 1314; 1315; 1325; 1326.
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.