EXACTAMIX 1200 Compounder, model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EM1200 DY Displ...

FDA Device Recall #Z-1963-2014 — Class II — June 10, 2014

Recall Summary

Recall Number Z-1963-2014
Classification Class II — Moderate risk
Date Initiated June 10, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Baxter Healthcare Corp.
Location Deerfield, IL
Product Type Devices
Quantity 1758

Product Description

EXACTAMIX 1200 Compounder, model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EM1200 DY Display, Replacement EM 1200 DY Display. EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX. Baxa Corporation. An automated pumping system that compounds multiple sterile ingredients into a finished solution in a single patient IV administration bag.

Reason for Recall

If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the original UI remaining in the tube which could be delivered into the next bag.

Distribution Pattern

US Nationwide

Lot / Code Information

model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX

Other Recalls from Baxter Healthcare Corp.

Recall # Classification Product Date
Z-0370-2017 Class II GAMBRO Capillary Dialyzer, Polyflux(R) Revaclea... Oct 12, 2016
Z-0670-2017 Class II sigma spectrum VOLUMETRIC INFUSION PUMP with Ma... Sep 28, 2016
Z-0671-2017 Class II Baxter SIGMA Spectrum INFUSION PUMP with Master... Sep 28, 2016
Z-0672-2017 Class II V6 Rear Case Assembly, Product Code 35701, a re... Sep 28, 2016
Z-2533-2016 Class II Synovis VASCU-GUARD Peripheral Vascular Patch i... Jun 24, 2016

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.