IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LK3PS1, SMN 10706293 (US...
FDA Device Recall #Z-1120-2017 — Class II — December 21, 2016
Recall Summary
| Recall Number | Z-1120-2017 |
| Classification | Class II — Moderate risk |
| Date Initiated | December 21, 2016 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Siemens Healthcare Diagnostics, Inc. |
| Location | Tarrytown, NY |
| Product Type | Devices |
| Quantity | 1379 kits (190 kits in U.S.A and 1189 Kits internationally) in total |
Product Description
IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LK3PS1, SMN 10706293 (US Kit Catalog) Product Usage: For in vitro diagnostic use with the IMMULITE¿ and IMMULITE 1000 Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients.
Reason for Recall
Negative bias results with the IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit was confirmed for two lots (D0119 and D0120). The investigation by Siemens has confirmed that patient samples exhibit an average negative bias of approximately -22% with a range of -16% to -27% for samples recovering from 0.08 to 16.2 ng/mL.
Distribution Pattern
Worldwide Distribution - US Nationwide and the countries of Denmark, Brazil, Croatia and Germany
Lot / Code Information
Lot Numbers: 1) D01196 - Expiration date 2016-12-31 2) D0120 - Expiration date 2017-03-31
Other Recalls from Siemens Healthcare Diagnostics, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1980-2026 | Class II | Dimension Creatinine Flex reagent cartridge. Ma... | Mar 18, 2026 |
| Z-1657-2026 | Class II | Atellica CH Enzymatic Hemoglobin A1c (A1c_E); ... | Feb 23, 2026 |
| Z-1484-2026 | Class II | Atellica CH Urine Albumin (UAlb). Material Numb... | Jan 20, 2026 |
| Z-0491-2026 | Class II | IMMULITE 2000 Oak Mix Specific Allergen; Sieme... | Oct 6, 2025 |
| Z-0492-2026 | Class II | 3gAllergy Specific IgE Universal Kit; Kit Comp... | Oct 6, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.