2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NO...

FDA Device Recall #Z-2738-2015 — Class II — July 31, 2015

Recall Summary

Recall Number Z-2738-2015
Classification Class II — Moderate risk
Date Initiated July 31, 2015
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Trilliant Surgical Ltd.
Location Houston, TX
Product Type Devices
Quantity 108 units

Product Description

2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plating System for trauma and reconstructive procedures of small bones.

Reason for Recall

Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.

Distribution Pattern

NY, OH, CA, SC, NJ, AZ, VA, IN, FL, MN, OK, AR, MO, WI, GA, SD, IA, IL, TX

Lot / Code Information

TSL002605

Other Recalls from Trilliant Surgical Ltd.

Recall # Classification Product Date
Z-2339-2018 Class II Ratcheting Cannulated Driver Handle, Catalog #:... Mar 27, 2018
Z-2338-2018 Class II Cannulated Driver Handle, Catalog #: 210-00-003... Mar 27, 2018
Z-0960-2017 Class II Tiger Screw Sterilization Tray Base Assembly Mo... Dec 8, 2016
Z-0609-2017 Class II Model 208-70-110, Long Thread Tiger Large Cannu... Sep 2, 2016
Z-2490-2015 Class II 7 Hole VL Gridlock Fibula Plate, Part # 300-60-... Aug 6, 2015

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.