NM/CT 860 Product Usage: The GE NM/CT 860 system is intended for general Nuclear Medicine...
FDA Device Recall #Z-1263-2019 — Class II — March 20, 2019
Recall Summary
| Recall Number | Z-1263-2019 |
| Classification | Class II — Moderate risk |
| Date Initiated | March 20, 2019 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | GE Healthcare, LLC |
| Location | Waukesha, WI |
| Product Type | Devices |
| Quantity | 6 units |
Product Description
NM/CT 860 Product Usage: The GE NM/CT 860 system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body and vascular X-ray Computed Tomography applications.
Reason for Recall
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Distribution Pattern
US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
Lot / Code Information
Mfg. Lot or Serial # System ID Model, Catalogue or Code # 860Y62005 SHIPPED (GON 4736850) H3908AD 850X61003 401793860NM H3908AD 860X62001 E6843969 H3908AE 870X64005 HC4387NU05 H3100AS 860Y62004 083026212959219 H3908AD 860X62002 PL0399NM01 H3908AD
Other Recalls from GE Healthcare, LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1566-2022 | Class II | (1) GE Centricity Universal Viewer 6.0, Model 2... | Jul 15, 2022 |
| Z-1430-2022 | Class II | SIGNA Premier magnetic resonance scanner, model... | Jun 24, 2022 |
| Z-1466-2022 | Class II | Centricity PACS Software Version 7.0 SP0.0.4.7 | Jun 14, 2022 |
| Z-1333-2022 | Class II | GE Centricity Universal Viewer Zero Footprint. ... | Jun 8, 2022 |
| Z-1511-2022 | Class II | Service manuals for the CT, PET, or NM table as... | Jun 2, 2022 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.