Surface Applicator Set with Leipzig-style Cone. Brachytherapy applicator set.
FDA Device Recall #Z-1829-2015 — Class II — May 18, 2015
Recall Summary
| Recall Number | Z-1829-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 18, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Varian Medical Systems Inc |
| Location | Charlottesville, VA |
| Product Type | Devices |
| Quantity | 46 |
Product Description
Surface Applicator Set with Leipzig-style Cone. Brachytherapy applicator set.
Reason for Recall
Varian has discovered that there is a discrepancy in the absolute dose rate given with the Leipzig-style Surface Applicator's Instructions for Use (IFU): Dose Characterization GM11010080 2012-09-06. The actual dose rate of the applicator is approximately 14% higher than the rate published within the IFU. Use of the dose rate as provided by the IFU without modification or independent confirmat
Distribution Pattern
Worldwide Distribution.
Lot / Code Information
JAQ
Other Recalls from Varian Medical Systems Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1849-2015 | Class II | The VariSource iX series afterloader systems ar... | May 18, 2015 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.