Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the ...

FDA Device Recall #Z-1382-2022 — Class II — June 2, 2022

Recall Summary

Recall Number Z-1382-2022
Classification Class II — Moderate risk
Date Initiated June 2, 2022
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Breas Medical, Inc.
Location North Billerica, MA
Product Type Devices
Quantity 1,113 units

Product Description

Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation Model Number: 230000 ¿ If the recalled product is a software, include version number: firmware versions 5.0.5 or earlier

Reason for Recall

A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could potentially be Permanent impairment or life threatening if medical intervention is not obtained

Distribution Pattern

Nationwide

Lot / Code Information

Unique Device Identifier (UDI): UDI 07321822300004. Serial number ranges in scope are: D******, K******, M01**** to M040124, where asterisk (*) denotes a digit from 0 to 9.

Other Recalls from Breas Medical, Inc.

Recall # Classification Product Date
Z-2690-2024 Class I Vivo 45 LS, pressure and volume ventilator capa... Jul 25, 2024
Z-0405-2024 Class II with Notch 22M/15F sold as part of Breas Medica... Oct 13, 2023
Z-0943-2022 Class II Breas Ventilator Trolley with the Vivo 50/60/65... Mar 21, 2022
Z-1352-2020 Class II Protective Cover Vivo 55/65 (Used for Vivo 50/6... Jan 28, 2020

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.