The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative...
FDA Device Recall #Z-1215-2015 — Class II — January 16, 2015
Recall Summary
| Recall Number | Z-1215-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | January 16, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Brainlab AG |
| Location | Feldkirchen, N/A |
| Product Type | Devices |
| Quantity | 114 offset cup impactors |
Product Description
The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive orthopedic surgery. (a stereotaxic instrument)
Reason for Recall
The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.
Distribution Pattern
Worldwide Distribution - US: Nationwide (AZ, CA, ID, IL, MD, MI, MN, OH, OK, OR, TN, WA, WI, WV); Australia, Austria, Bahrain, Belgium, Germany, Hong Kong, India, Italy, Japan, Malaysia, Russian Federation, Saudi Arabia, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom
Lot / Code Information
Model 52856 BL CURVED CUP INSERTER UNIVERSAL SHAFT; and MODEL 52856A BL OFFSET CUP IMPACTOR UNIVERSAL SHAFT, all serial/lot numbers distributed prior to January 13, 2015.
Other Recalls from Brainlab AG
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1392-2025 | Class II | Origin Data Management software versions 3.1.0,... | Feb 10, 2025 |
| Z-1292-2023 | Class II | Cirq Arm System 2.0 | Mar 2, 2023 |
| Z-1259-2023 | Class II | Brainlab ExacTrac Dynamic software, Medical Cha... | Feb 14, 2023 |
| Z-0733-2022 | Class II | ExacTrac Dynamic software model numbers: 209... | Feb 25, 2022 |
| Z-1846-2021 | Class II | ExacTrac Dynamic software, Model 20910-01B ETD ... | May 14, 2021 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.