Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno
FDA Recall #Z-0656-2021 — Class II — November 30, 2020
Product Description
Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno
Reason for Recall
When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits.
Recalling Firm
Siemens Medical Solutions USA, Inc — Malvern, PA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
5602
Distribution
Worldwide-US Nationwide
Code Information
Model number: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Pheno- 10849000 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355
Status
Ongoing
Voluntary / Mandated
FDA Mandated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.