Baxter Amia Automated PD systems are used in the treatment of adult renal failure patients to per...
FDA Device Recall #Z-0932-2015 — Class II — April 24, 2014
Recall Summary
| Recall Number | Z-0932-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | April 24, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Baxter Healthcare Corp. |
| Location | Deerfield, IL |
| Product Type | Devices |
| Quantity | 55 units |
Product Description
Baxter Amia Automated PD systems are used in the treatment of adult renal failure patients to perform automated peritoneal dialysis.
Reason for Recall
System error 01779 is produced when the battery cannot be charged due to specific voltage differences between the battery and the charger.
Distribution Pattern
Nationwide Distribution including GA, MA, NC, NY, PA, and WA.
Lot / Code Information
Product Code: 5C9310
Other Recalls from Baxter Healthcare Corp.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0370-2017 | Class II | GAMBRO Capillary Dialyzer, Polyflux(R) Revaclea... | Oct 12, 2016 |
| Z-0670-2017 | Class II | sigma spectrum VOLUMETRIC INFUSION PUMP with Ma... | Sep 28, 2016 |
| Z-0671-2017 | Class II | Baxter SIGMA Spectrum INFUSION PUMP with Master... | Sep 28, 2016 |
| Z-0672-2017 | Class II | V6 Rear Case Assembly, Product Code 35701, a re... | Sep 28, 2016 |
| Z-2533-2016 | Class II | Synovis VASCU-GUARD Peripheral Vascular Patch i... | Jun 24, 2016 |
Frequently Asked Questions
Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.