Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.

FDA Recall #Z-0810-2013 — Class II — December 17, 2012

Recall #Z-0810-2013 Date: December 17, 2012 Classification: Class II Status: Terminated

Product Description

Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.

Reason for Recall

Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.

Recalling Firm

Mizuho OSI — Union City, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

869 devices

Distribution

USA Distribution -- Worldwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and WY., and the countries of Australia, Canada, Egypt, Finland, France, Germany, China, Israel, Italy, Japan, Netherlands, New Zealand, P.R. China, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Taiwan R.O.C., UAE and UK.

Code Information

Table model 5855, sub assembly 5855-901 Orthopedic Table Spar, manufactured between 1996 and 2007.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated