Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 7...

FDA Recall #Z-2478-2025 — Class II — August 4, 2025

Recall #Z-2478-2025 Date: August 4, 2025 Classification: Class II Status: Ongoing

Product Description

Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.

Reason for Recall

Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

Recalling Firm

WASATCH PHOTONICS — Morrisville, NC

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

192

Distribution

U.S.

Code Information

all codes included in the model numbers.

Status

Ongoing

Voluntary / Mandated

FDA Mandated