Model No. CYF-V2R, VISERA Cysto-Nephro Videoscope
FDA Recall #Z-0692-2024 — Class II — December 4, 2023
Product Description
Model No. CYF-V2R, VISERA Cysto-Nephro Videoscope
Reason for Recall
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Recalling Firm
Shirakawa Olympus Co., Ltd. — Nishishirakawa-Gun
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
608 US; 1264 OUS
Distribution
US Nationwide distribution.
Code Information
UDI-DI: 04953170339523 04953170411212 All serial numbers
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated