Covidien Tri-Staple 2.0 Reinforced Intelligent Reload 60 mm Medium/Thick-For use with Signia stap...
FDA Device Recall #Z-1419-2021 — Class II — March 23, 2021
Recall Summary
| Recall Number | Z-1419-2021 |
| Classification | Class II — Moderate risk |
| Date Initiated | March 23, 2021 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Covidien Llc |
| Location | Mansfield, MA |
| Product Type | Devices |
| Quantity | 222 units |
Product Description
Covidien Tri-Staple 2.0 Reinforced Intelligent Reload 60 mm Medium/Thick-For use with Signia stapling system. Staple line reinforcement has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas. Product Number: SIGTRSB60AMT
Reason for Recall
Potential for the pouch to contain an extra laminate layer, creating the potential for the device sterility to be compromised. The inclusion of an additional laminate layer in the package may also impact complete de-gassing of the sterilization gas, Ethylene Oxide and may result in a potentially increased risk for patient infection and/or potential exposure to Ethylene Oxide retained in the pouch
Distribution Pattern
AL, AR, FL, IN, LA, MA , MN, MO, NJ, NY, OR, PA, TX
Lot / Code Information
Lot Number: N0M0801RY Exp. Date: ¿11/30/2023 UDI: A8845217176201 20884521717623
Other Recalls from Covidien Llc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1557-2026 | Class II | HealthCast "Vital Sync" Remote Patient Monitori... | Jan 30, 2026 |
| Z-1775-2022 | Class II | Puritan Bennett 560 Ventilator, PB560, CFN 4096600 | Aug 19, 2022 |
| Z-1318-2022 | Class I | Palindrome SI Chronic Catheter, Sterile, Single... | Jun 8, 2022 |
| Z-1317-2022 | Class I | Palindrome H Chronic Catheter, Sterile, Single ... | Jun 8, 2022 |
| Z-1319-2022 | Class I | Palindrome HSI Chronic Catheter, Sterile, Singl... | Jun 8, 2022 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.