BRAVO PH CAPS DELIVERY DEV5-PK, FGS-0312/FGS-0313 - Product Usage: intended to be used for gastro...
FDA Device Recall #Z-1335-2021 — Class II — February 26, 2021
Recall Summary
| Recall Number | Z-1335-2021 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 26, 2021 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Covidien Llc |
| Location | Mansfield, MA |
| Product Type | Devices |
| Quantity | 62551 |
Product Description
BRAVO PH CAPS DELIVERY DEV5-PK, FGS-0312/FGS-0313 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age.
Reason for Recall
The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.
Distribution Pattern
Worldwide distribution - US Nationwide distribution.
Lot / Code Information
49349Q 49350Q 49351Q 49352Q 49799Q 49800Q 49801Q 49802Q 49809Q 49849Q 49850Q 49851Q 50392Q 50393Q 50492Q 50493Q 50494Q 50495Q 50496Q 50497Q 50615Q 50616Q 50868Q 50869Q 50870Q 50871Q 50872Q 50873Q 50874Q 50875Q 50876Q 50877Q 50878Q 50879Q 50880Q 50881Q
Other Recalls from Covidien Llc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1557-2026 | Class II | HealthCast "Vital Sync" Remote Patient Monitori... | Jan 30, 2026 |
| Z-1775-2022 | Class II | Puritan Bennett 560 Ventilator, PB560, CFN 4096600 | Aug 19, 2022 |
| Z-1318-2022 | Class I | Palindrome SI Chronic Catheter, Sterile, Single... | Jun 8, 2022 |
| Z-1319-2022 | Class I | Palindrome HSI Chronic Catheter, Sterile, Singl... | Jun 8, 2022 |
| Z-1315-2022 | Class I | Palindrome Precision HSI Chronic Catheter, Ster... | Jun 8, 2022 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.