iVue and iVue 500 (now called iScan) OCT systems with software version 2016.0.0.127 (USA only) an...

FDA Device Recall #Z-1498-2016 — Class II — March 16, 2016

Recall Summary

Recall Number Z-1498-2016
Classification Class II — Moderate risk
Date Initiated March 16, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Optovue, Inc.
Location Fremont, CA
Product Type Devices
Quantity 40

Product Description

iVue and iVue 500 (now called iScan) OCT systems with software version 2016.0.0.127 (USA only) and 2016.1.0.127 (out of US) Opthalmic: iVue is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber the eye.

Reason for Recall

A software anomaly was detected that affects Optovue iVue and iVue 500 with Normative Database that may display images of the same eye rather than right and left if user scrolls on the map.

Distribution Pattern

PA, CA, IL, HI. MO, CO, GA, ME, IN, AZ, WV, VA, FL, AR, UT, NE, IN, NV, LA, TN, KS, OK. Foreign distribution forthcoming

Lot / Code Information

US and foreign (not yet provided) Software version 2016.0.0.127: Serial numbers: 21908, 26507, 26496, 26558, 22738, 26139, 26442, 25720, 25079, 25142, 23298, 21990, 23461, 24199, 25843, 26572, 26550, 22971, 26495, 22652, 26550, 22971, 26495, 22652, 25389, 23309, 23760, 21732-1, 23527, 20923, 21600, 21414, 20271, 23153, 22875, 23211. Software version 2016.1.0.127; Serial numbers forthcoming.

Other Recalls from Optovue, Inc.

Recall # Classification Product Date
Z-0009-2018 Class II iFusion, Device Listing Number D203271. Prod... Jul 27, 2017
Z-0008-2018 Class II iVue with Normative Database, Device Listing Nu... Jul 27, 2017
Z-0010-2018 Class II iVue 500 with iScan, D216606. Product non-c... Jul 27, 2017
Z-1441-2013 Class III Optovue iVue with Normative Database (NDB) with... May 14, 2013

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.