EasyOne Pro LAB VOS Respiratory Analysis System Pulmonary Function Testing Device, Model 3100-1
FDA Device Recall #Z-0597-2020 — Class II — August 15, 2017
Recall Summary
| Recall Number | Z-0597-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 15, 2017 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | NDDd Medical Technologies, Inc. |
| Location | Andover, MA |
| Product Type | Devices |
| Quantity | 37 total devices |
Product Description
EasyOne Pro LAB VOS Respiratory Analysis System Pulmonary Function Testing Device, Model 3100-1
Reason for Recall
The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lower electromagnetic shielding provided by the cable harness, the CO/CO2 sensor could have been influenced by high frequencies emitted by very close devices.
Distribution Pattern
The products were distributed to the following US states: KY, MO, NY, OH, PA
Lot / Code Information
Serial Numbers: 650074 650085 650075 650086 650076 650087 650079 650088 650080 650089 650081 650090 650082 650091 650083 650092 650094 650095 650096 650098 650099 650100 650104 650055 650058 650062 650066 650067 650069 650070 650073
Other Recalls from NDDd Medical Technologies, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0596-2020 | Class II | EasyOne Pro Respiratory Analysis System Pulmona... | Aug 15, 2017 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.