HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or d...
FDA Device Recall #Z-1329-2024 — Class II — March 11, 2024
Recall Summary
| Recall Number | Z-1329-2024 |
| Classification | Class II — Moderate risk |
| Date Initiated | March 11, 2024 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Baxter Healthcare Corporation |
| Location | Deerfield, IL |
| Product Type | Devices |
| Quantity | 80 units |
Product Description
HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery
Reason for Recall
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.
Distribution Pattern
US Nationwide distribution in the state of Florida.
Lot / Code Information
123510000293, 123510000294, 123510000295, 123510000296, 123510000298, 123510000299, 123510000300, 123510000764, 123510000765, 123510000766, 123510000767, 123510000768, 123510000769, 123510000770, 123510000771, 123510000772, 123510000773, 123510000800, 123510000801, 123510000802, 123510000803, 123510000804, 123510000805, 123510000806, 123510000807, 123510000808, 123510000809, 123510000297, 123350001033, 123350001034, 123350001035, 123350001036, 123350001040, 123350001041, 123350001042, 123360000294, 123360000295, 123360000297, 123360000298, 123360000299, 123360000300, 123360000301, 123360000302, 123360000303, 123360000304, 123360000733, 123360000734, 123360000735, 123360000736, 123360000737, 123360000738, 123360000739, 123360000740, 123360001972, 123360001973, 123360001974, 123360001975, 123360001976, 123360001977, 123360001978, 123360001979, 123360001980, 123360001981, 123350000968, 123350000969, 123350000970, 123350000971, 123350000972, 123350000973, 123350000974, 123350000975, 123350000987, 123350000988, 123350000989, 123350000990, 123350000991, 123350000992, 123350000993, 123350000994, 123350001032
Other Recalls from Baxter Healthcare Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1573-2026 | Class II | Brand Name: Welch Allyn, Inc. Product Name: We... | Feb 18, 2026 |
| Z-1368-2026 | Class II | Brand Name: Spectrum IQ Product Name: Infusion... | Jan 13, 2026 |
| Z-1121-2026 | Class II | SIGMA Spectrum Infusion Pump, Model/Catalog Num... | Nov 28, 2025 |
| Z-0847-2026 | Class II | Welch Allyn INF BAG, Thigh 1-Tube, Model/Catalo... | Oct 28, 2025 |
| Z-0233-2026 | Class II | Oral Probe (Product code 02893-000), accessory ... | Sep 17, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.