Pentax Medical C2 CryoBalloon Pear Focal Catheter, Model FG 1024, Sterile, RX. Product Usage: ...

FDA Recall #Z-1326-2019 — Class II — December 3, 2018

Recall #Z-1326-2019 Date: December 3, 2018 Classification: Class II Status: Terminated

Product Description

Pentax Medical C2 CryoBalloon Pear Focal Catheter, Model FG 1024, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus. The system includes the following components: (1) CryoBalloon Catheter connects to the Controller, which controls the operation of the Catheter such as diffuser (sprayer) positioning and ablation (nitrous oxide release); (2) CryoBalloon Controller contains the cartridge heater and cryogen delivery valve, which is controlled with the foot pedal; (3) Foot Pedal; and (4) CryoBalloon Cartridge containing 36 grams of nitrous oxide.

Reason for Recall

Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.

Recalling Firm

PENTAX of America Inc — Redwood City, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

68 devices

Distribution

Distribution was made to CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, and WA. There was no government/military distribution. Foreign distribution was made to Canada, Germany, and the Netherlands.

Code Information

Lot/Serial numbers: 06152018-01, 07172018-02, 07022018-04, 08212018-01, and 08102018-03.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.