MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)

FDA Device Recall #Z-1574-2015 — Class II — April 17, 2015

Recall Summary

Recall Number Z-1574-2015
Classification Class II — Moderate risk
Date Initiated April 17, 2015
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Location HIROSHIMA, N/A
Product Type Devices
Quantity 11 units total

Product Description

MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)

Reason for Recall

The operator console allows users to deliver therapeutic radiation to patients even though a specific communication error relevant to imaging conditions has occurred. In addition, the treatment record (the delivered radiation record) cannot be saved.

Distribution Pattern

Worldwide Distribution - US including FL, NY, OH, & TX, **Internationally - Belgium, Italy, Korea, France & Germany.

Lot / Code Information

Serial Numbers: 201902, 201903, 202903, 202905, 203901, 203902, 203903, 203905, 203906, 203919 & 203924

Other Recalls from MITSUBISHI HEAVY INDUSTRIES, LTD., HI...

Recall # Classification Product Date
Z-1308-2016 Class II VERO MHI-TM2000 Linear Accelerator System --- ... Mar 7, 2016
Z-1244-2016 Class II MHI-TM2000 Linear Accelerator System (Software ... Dec 15, 2015
Z-0821-2015 Class II VERO Linear Accelerator System, Model Number MH... Dec 2, 2014
Z-2653-2014 Class II VERO Linear Accelerator System, Model Number M... Jul 5, 2014
Z-1931-2013 Class II MHI-TM2000 Linear Accelerator System, Software ... Jun 21, 2013

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.