Neodent Easypack Helix GM Acqua 3.5X13/Smart 4.5X2.5 Article Number: 138.107
FDA Device Recall #Z-0513-2022 — Class II — November 30, 2021
Recall Summary
| Recall Number | Z-0513-2022 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 30, 2021 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Straumann USA LLC |
| Location | Andover, MA |
| Product Type | Devices |
| Quantity | 1 unit |
Product Description
Neodent Easypack Helix GM Acqua 3.5X13/Smart 4.5X2.5 Article Number: 138.107
Reason for Recall
Mislabeled: Carton box item 138.089- Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 has an implant with measure of 3.5x13mm; and Carton Box (138.107- Easypack Helix GM Acqua has a 3.5x8mm implant instead of a 3.5x 13mm implant
Distribution Pattern
US Nationwide Distribution in the states of FL, IL, OH
Lot / Code Information
Lot GNV20: UDI:7899878052906
Other Recalls from Straumann USA LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1994-2026 | Class II | Straumann n!ce PMMA Full-arch Restoration. Scre... | Mar 27, 2026 |
| Z-1995-2026 | Class II | Straumann n!ce Zr, HT, Full-arch Restoration. A... | Mar 27, 2026 |
| Z-1632-2026 | Class II | Custom Abutment Ti, for Straumann RC, incl. TAN... | Mar 17, 2026 |
| Z-1633-2026 | Class II | Custom Abutment AS Ti, for Straumann RC, incl. ... | Mar 17, 2026 |
| Z-1337-2026 | Class III | WB Impression Post Closed Tray incl. guide scre... | Jan 6, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.