STERRAD NX, Product Code: 10033 The STERRAD Steri lization System is a low temperature general...
FDA Device Recall #Z-0403-2014 — Class II — November 11, 2013
Recall Summary
| Recall Number | Z-0403-2014 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 11, 2013 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Advanced Sterilization Products |
| Location | Irvine, CA |
| Product Type | Devices |
| Quantity | Total 4,381 units |
Product Description
STERRAD NX, Product Code: 10033 The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
Reason for Recall
Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.
Distribution Pattern
Nationwide Distribution
Lot / Code Information
Please refer to consignee list
Other Recalls from Advanced Sterilization Products
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1197-2017 | Class II | STERRAD NX Sterilization System, Product Code: ... | Jan 9, 2017 |
| Z-0965-2017 | Class II | EVOTECH Endoscope Cleaner and Reprocessor Syste... | Sep 27, 2016 |
| Z-2774-2015 | Class II | CIDEX OPA Solution, P/N 20390 for use as a high... | Aug 10, 2015 |
| Z-2459-2015 | Class II | CIDEX Activated Dialdehyde Solution, 4 bottles/... | Jul 27, 2015 |
| Z-2375-2015 | Class II | Wall Chart for STERRAD 100NX System, P/N 10104.... | Jul 17, 2015 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.